Performance Indicators

Blank absorbance: The blank absorbance measured at A600nm should be ≤0.05.
Accuracy: Using quality control products for determination, the average of the measured results should be within the target value range of the quality control product.
Precision: Repeat the determination of the same concentration sample or quality control product 10 times. The coefficient of variation (CV%) of the measured values should not be greater than 10%.
Linearity: Within the range of [0.104,2.59] mmol/L ([4.0,100.0] mg/dL), the linear correlation coefficient r ≥ 0.990.
Sensitivity: When the sample concentration is 2.59 mmol/L (100.0 mg/dL), the change in absorbance is between 0.0500 and 0.2500.