Performance Indicators
Accuracy: Determined using quality control products, the average of the measurement results should be within the target value range of the quality control products.
Precision: a、Repeatability: When the same quality control or serum sample is measured repeatedly 10 times, the coefficient of variation (CV%) of the measured values should not exceed 10%.
b、Inter-batch difference: The inter-batch difference of three batches of reagent kits randomly selected should be ≤10%.
Linearity: Within the range of [0.5,130] ng/mL, the linear correlation coefficient r ≥ 0.990. Within the range of [0.5,50] ng/mL, the linear deviation should not exceed ±5 ng/mL. Within the range of (50,130] ng/mL, the measurement deviation should not exceed ±10%.
Sensitivity: When the sample concentration is 20 ng/mL, the change in absorbance is between 0.0100 and 0.1000.