Performance Indicators

Accuracy: Determined using quality control products. The average of the measured results should fall within the target range of the quality control products.
Precision: a, Repeatability: When the same serum sample is measured repeatedly 10 times, the coefficient of variation (CV%) of the measured values should not exceed 10%.
b, Inter-batch difference: The inter-batch difference of three randomly selected reagent kits should be ≤10%.
Linearity: Within the range of [10,200] mg/dL, the linear correlation coefficient r²≥0.99. The linear deviation within the range of [10,20] mg/dL should not exceed ±2 mg/dL, and the measurement deviation within the range of (20,200] mg/dL should not exceed ±10%.
Sensitivity: When the sample concentration is 60 mg/dL, the absolute value of the rate of change in absorbance per minute is between 0.01 and 0.10.

Get Free Consultation

Note: We will contact you within one working day. Please pay attention to your email.

Submit Now

Recommended Products

Your Health has standard and complete production conditions and management levels, ensuring product quality output.