Ischemia Modified Albumin (IMA) Assay Kit
performance index
Accuracy: The quality control product is used for measurement. The average value of the measurement results should be within the range of quality control target values.
Precision: a, repeatability: repeated determination 10 times using the same quality control or serum sample, and the coefficient of variation(CV %) of the measured value should ≤ 5 %.
B, interbatch difference: The interbatch difference for randomly selecting three batch kits should ≤ 10 %.
Linear: in[ 20,100] In the U/mL range, the linear correlation coefficient R ≥ 0.990, in[ 20, 50] The linear deviation in the U/mL range should not exceed ± 5U/mL in(50,100) The determination deviation within the U / mL range shall not exceed ± 10 %.
Sensitivity: When the sample concentration is 60U/mL, the absorbance varies from 0.1500 to 0.6500.